The project aims to experiment the Cone Beam Breast Computed Tomography technique using a standard digital mammography system. The work is focused on the definition of a protocol of quality measurements for the pre-clinical evaluation of the machine.
The paper is developed in two parts. The first is specifically concerned with the methods used to define the image quality and dosimetry aspects specific for digital mammography devices. A complete characterization of the system has been performed according to the applicable IEC standards to assure the performances of the equipment and define the quality levels. Due to the lack of a quality control protocol dedicated to CBBCT mammography scanner, a new equivalent test procedure has been proposed.
The second part of the paper is focused on the evaluation, through quantitative and visual analyzes, of the CBCT exam feasibility in the hardware and software conditions currently proposed by IMS Giotto. The prototype was in fact developed differing from the technical choices of competing companies and developed for a different intended use. The main difference with respect to the existing breast CT scanners is the possibility of performing on the same system the CBBCT scanning but also all the mammographic techniques. In this thesis, we aim to assess whether, in the current setup, considering a dosimetric range very close to that used in the clinic, the tests produce results that can be considered acceptable or at least indicative of the feasibility of the entire project from a commercial point of view. For this purpose, the final reconstruction images, obtained by two previously developed software, are analyzed.
Abstract
The project aims to experiment the Cone Beam Breast Computed Tomography technique using a standard digital mammography system. The work is focused on the definition of a protocol of quality measurements for the pre-clinical evaluation of the machine.
The paper is developed in two parts. The first is specifically concerned with the methods used to define the image quality and dosimetry aspects specific for digital mammography devices. A complete characterization of the system has been performed according to the applicable IEC standards to assure the performances of the equipment and define the quality levels. Due to the lack of a quality control protocol dedicated to CBBCT mammography scanner, a new equivalent test procedure has been proposed.
The second part of the paper is focused on the evaluation, through quantitative and visual analyzes, of the CBCT exam feasibility in the hardware and software conditions currently proposed by IMS Giotto. The prototype was in fact developed differing from the technical choices of competing companies and developed for a different intended use. The main difference with respect to the existing breast CT scanners is the possibility of performing on the same system the CBBCT scanning but also all the mammographic techniques. In this thesis, we aim to assess whether, in the current setup, considering a dosimetric range very close to that used in the clinic, the tests produce results that can be considered acceptable or at least indicative of the feasibility of the entire project from a commercial point of view. For this purpose, the final reconstruction images, obtained by two previously developed software, are analyzed.
Tipologia del documento
Tesi di laurea
(Laurea magistrale)
Autore della tesi
Signorile, Emma
Relatore della tesi
Correlatore della tesi
Scuola
Corso di studio
Indirizzo
Applied Physics
Ordinamento Cds
DM270
Parole chiave
cone beam breast computed tomography,mammography,breast cancer,diagnosis,characterization,reconstruction
Data di discussione della Tesi
28 Ottobre 2022
URI
Altri metadati
Tipologia del documento
Tesi di laurea
(NON SPECIFICATO)
Autore della tesi
Signorile, Emma
Relatore della tesi
Correlatore della tesi
Scuola
Corso di studio
Indirizzo
Applied Physics
Ordinamento Cds
DM270
Parole chiave
cone beam breast computed tomography,mammography,breast cancer,diagnosis,characterization,reconstruction
Data di discussione della Tesi
28 Ottobre 2022
URI
Statistica sui download
Gestione del documento: