Bonali, Elena
(2026)
Commissioning of the neutron veto of the XENONnT experiment.
[Laurea magistrale], Università di Bologna, Corso di Studio in
Physics [LM-DM270]
Documenti full-text disponibili:
![[thumbnail of Thesis]](https://amslaurea.unibo.it/style/images/fileicons/application_pdf.png) |
Documento PDF (Thesis)
Disponibile con Licenza: Salvo eventuali più ampie autorizzazioni dell'autore, la tesi può essere liberamente consultata e può essere effettuato il salvataggio e la stampa di una copia per fini strettamente personali di studio, di ricerca e di insegnamento, con espresso divieto di qualunque utilizzo direttamente o indirettamente commerciale. Ogni altro diritto sul materiale è riservato
Download (22MB)
|
Abstract
The XENONnT experiment at the Laboratori Nazionali del Gran Sasso utilizes a dual-phase liquid xenon Time Projection Chamber for Dark Matter WIMP search. Neutron-induced single scatters are a critical background, as they can mimic WIMP signals. To mitigate this, XENONnT employs a gadolinium-doped water Cherenkov veto—the Neutron Veto (NV).
This thesis presents the recommissioning of the NV after a maintenance carried out in 2025. Operations progressed from initial validation with pure water to doping with gadolinium sulfate (GdSO), reaching a concentration of 2500 ppm, 5 times higher than in the pre-upgrade science run.
Systematic measurements of photomultiplier tube parameters confirmed stable performance. Laser calibrations and event-timing observables showed well-preserved optical properties after GdSO loading, degrading by less than 5% relative to pure-water. Backgrounds were characterized by rate evolution and sources.
Calibration data with an americium-beryllium (AmBe) source validated the detector response, demonstrating a reduced neutron capture time of 25 μs. This allows the veto coincidence window to be reduced from 250 to 100 μs, decreasing the live-time loss by a factor of 2.5.
The neutron tagging efficiency—the fraction of single-scatter nuclear recoils in the WIMP region tagged by neutron capture in the NV—is the key parameter for evaluating background rejection capability. Since dedicated AmBe data for measurement at the current concentration are not yet available, this work provides an estimate based on Monte Carlo simulations and previous measurements. For the current concentration, the neutron tagging efficiency is (89.5 ± 1.4)%, a substantial improvement over the configurations used previously. An efficiency of (92.0 ± 1.7)% is projected for the future target concentration of 5000 ppm.
Overall, this work demonstrates the stable operation and improved performance of the NV, establishing its readiness for the upcoming XENONnT Dark Matter search.
Abstract
The XENONnT experiment at the Laboratori Nazionali del Gran Sasso utilizes a dual-phase liquid xenon Time Projection Chamber for Dark Matter WIMP search. Neutron-induced single scatters are a critical background, as they can mimic WIMP signals. To mitigate this, XENONnT employs a gadolinium-doped water Cherenkov veto—the Neutron Veto (NV).
This thesis presents the recommissioning of the NV after a maintenance carried out in 2025. Operations progressed from initial validation with pure water to doping with gadolinium sulfate (GdSO), reaching a concentration of 2500 ppm, 5 times higher than in the pre-upgrade science run.
Systematic measurements of photomultiplier tube parameters confirmed stable performance. Laser calibrations and event-timing observables showed well-preserved optical properties after GdSO loading, degrading by less than 5% relative to pure-water. Backgrounds were characterized by rate evolution and sources.
Calibration data with an americium-beryllium (AmBe) source validated the detector response, demonstrating a reduced neutron capture time of 25 μs. This allows the veto coincidence window to be reduced from 250 to 100 μs, decreasing the live-time loss by a factor of 2.5.
The neutron tagging efficiency—the fraction of single-scatter nuclear recoils in the WIMP region tagged by neutron capture in the NV—is the key parameter for evaluating background rejection capability. Since dedicated AmBe data for measurement at the current concentration are not yet available, this work provides an estimate based on Monte Carlo simulations and previous measurements. For the current concentration, the neutron tagging efficiency is (89.5 ± 1.4)%, a substantial improvement over the configurations used previously. An efficiency of (92.0 ± 1.7)% is projected for the future target concentration of 5000 ppm.
Overall, this work demonstrates the stable operation and improved performance of the NV, establishing its readiness for the upcoming XENONnT Dark Matter search.
Tipologia del documento
Tesi di laurea
(Laurea magistrale)
Autore della tesi
Bonali, Elena
Relatore della tesi
Correlatore della tesi
Scuola
Corso di studio
Indirizzo
NUCLEAR AND SUBNUCLEAR PHYSICS
Ordinamento Cds
DM270
Parole chiave
Dark Matter,Neutron Veto,Gd-doped water,XENON experiment,WIMP,Direct detection,Neutron capture,Cherenkov
Data di discussione della Tesi
29 Ottobre 2026
URI
Altri metadati
Tipologia del documento
Tesi di laurea
(NON SPECIFICATO)
Autore della tesi
Bonali, Elena
Relatore della tesi
Correlatore della tesi
Scuola
Corso di studio
Indirizzo
NUCLEAR AND SUBNUCLEAR PHYSICS
Ordinamento Cds
DM270
Parole chiave
Dark Matter,Neutron Veto,Gd-doped water,XENON experiment,WIMP,Direct detection,Neutron capture,Cherenkov
Data di discussione della Tesi
29 Ottobre 2026
URI
Statistica sui download
Gestione del documento: